ASTM E1086 PDF

Buy ASTM E TEST METHOD FOR ANALYSIS OF AUSTENITIC STAINLESS STEEL BY SPARK ATOMIC EMISSION. ASTM E Standard Test Method for Analysis of Austenitic Stainless Steel by Spark Atomic Emission Spectrometry. standard by ASTM. ASTM E – Standard Test Method for optical emission vacuum spectrometric analysis of stainless.

Author: Dounos Vokree
Country: Hungary
Language: English (Spanish)
Genre: Literature
Published (Last): 8 October 2017
Pages: 451
PDF File Size: 2.45 Mb
ePub File Size: 20.35 Mb
ISBN: 137-5-58294-633-7
Downloads: 94950
Price: Free* [*Free Regsitration Required]
Uploader: JoJojora

The agreement between the results obtained by chemical methods and those obtained by this test method is displayed in Table 3. Reagents and Materials 7. This procedure is suitable for manufacturing control and inspection testing.

ASTM E1086 – 14

Make the necessary corrections either by adjusting the controls on the adtm or by applying arithmetic corrections. Summary of Test Method 4. Work Item s – proposed revisions of this standard.

Any of these that will achieve analytical accuracy equivalent to that reported for this test method are acceptable. Automatic sequencing shall be provided to actuate the? These data were obtained by procedures outlined in Practice E The following mold types have been found to produce acceptable samples.

The ring is placed on a? The discharge is terminated at a predetermined level of accumulated radiant energy from the internal standard iron line or after a? The precision data for the three unknowns is shown in Table 2.

  LA VIDA EN LA TIERRA AUDESIRK 8 EDICION PDF

Astmm lines are listed in Table 1. Excitation and Exposure One or more of the reference materials must closely approximate the composition of atsm specimen. The samples must be sufficiently massive to prevent overheating during the discharge and of a similar metallurgical condition asm composition as the reference materials.

ASTM International – Standard References for ASTM E – 14

Department of Commerce, Gaithersburg, MD Originally approved in The technique of analyzing reference materials along with unknowns and performing the e1806 mathematical corrections may also be used to correct for interference effects and to compensate for errors resulting from instrument drift. At the end of the exposure period, the charge on each capacitor is measured, and displayed or recorded as a relative energy or concentration.

Active view current version of standard. Using reference materials supplied with the unknowns, in-house reference materials, NBS and British Certi? The results for each individual laboratory is on?

Flush, s Preburn, s Exposure, s 2 10 10 or 15 Argon Flow 0. The samples must be sufficiently massive to prevent overheating during the discharge and of a similar metallurgical condition and composition as the reference materials.

If you feel that your comments have not received a fair hearing you should make your views known to the ASTM Committee on Standards, at the address shown below. The stand shall provide an atmosphere of argon and may be water cooled.

  EL ACROSPORT EN LA ESCUELA PDF

This represents a total of 15 determinations. Interference corrections will not be necessary if: A minimum of three reference materials shall be used for each element. A superscript epsilon e indicates an editorial change since the last revision or reapproval. A number in parentheses indicates the year of last reapproval.

Equipment varies among laboratories. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. The samples must be sufficiently massive to prevent overheating during the discharge and of a similar metallurgical condition and composition as the reference materials.

Standardization will be done anytime veri? H Resistance, V Potential, V 7. Referenced Documents purchase separately The documents listed below are referenced within the subject standard but are not provided as part of the standard.

F d-c charged Inductance,? Precision and Bias 2 Follow the procedure for the concentration-ratio method in accordance with Practice E Calculation of Results